Small tip angle three‐dimensional tailored radiofrequency slab‐select pulse for reduced <i>B</i><sub>1</sub> inhomogeneity at 3 T
98
Zitationen
4
Autoren
2005
Jahr
Abstract
This work presents a small tip-angle 3D tailored RF slab-select pulse for reducing the B1 field inhomogeneity at 3T. The compensated slice profile was determined from a B1 inhomogeneity map. SNR improvement and degree of artifact reduction were evaluated in a NiCl2 doped phantom and human brains. The technique was found to reduce inhomogeneities as large as 30% of the peak image magnitude in all three spatial directions in the brain using a standard head coil.