Dies ist eine Übersichtsseite mit Metadaten zu dieser wissenschaftlichen Arbeit. Der vollständige Artikel ist beim Verlag verfügbar.
Correction for geometric distortion in echo planar images from B<i>0</i> field variations
1.440
Zitationen
2
Autoren
1995
Jahr
Abstract
A method is described for the correction of geometric distortions occurring in echo planar images. The geometric distortions are caused in large part by static magnetic field inhomogeneities, leading to pixel shifts, particularly in the phase encode direction. By characterizing the field inhomogeneities from a field map, the image can be unwarped so that accurate alignment to conventionally collected images can be made. The algorithm to perform the unwarping is described, and results from echo planar images collected at 1.5 and 4 Tesla are shown.
Ähnliche Arbeiten
Advances in functional and structural MR image analysis and implementation as FSL
2004 · 13.972 Zit.
A default mode of brain function
2001 · 12.298 Zit.
FSL
2011 · 11.554 Zit.
Improved Optimization for the Robust and Accurate Linear Registration and Motion Correction of Brain Images
2002 · 10.587 Zit.
Functional connectivity in the motor cortex of resting human brain using echo‐planar mri
1995 · 9.999 Zit.