Comparison of silver and molybdenum microfocus X-ray sources for single-crystal structure determination
4.621
Zitationen
4
Autoren
2014
Jahr
Abstract
The quality of diffraction data obtained using silver and molybdenum microsources has been compared for six model compounds with a wide range of absorption factors. The experiments were performed on two 30 W air-cooled Incoatec IµS microfocus sources with multilayer optics mounted on a Bruker D8 goniometer with a SMART APEX II CCD detector. All data were analysed, processed and refined using standard Bruker software. The results show that Ag <i>K</i>α radiation can be beneficial when heavy elements are involved. A numerical absorption correction based on the positions and indices of the crystal faces is shown to be of limited use for the highly focused microsource beams, presumably because the assumption that the crystal is completely bathed in a (top-hat profile) beam of uniform intensity is no longer valid. Fortunately the empirical corrections implemented in <i>SADABS</i>, although originally intended as a correction for absorption, also correct rather well for the variations in the effective volume of the crystal irradiated. In three of the cases studied (two Ag and one Mo) the final <i>SHELXL R</i>1 against all data after application of empirical corrections implemented in <i>SADABS</i> was below 1%. Since such corrections are designed to optimize the agreement of the intensities of equivalent reflections with different paths through the crystal but the same Bragg 2θ angles, a further correction is required for the 2θ dependence of the absorption. For this, <i>SADABS</i> uses the transmission factor of a spherical crystal with a user-defined value of μ<i>r</i> (where μ is the linear absorption coefficient and <i>r</i> is the effective radius of the crystal); the best results are obtained when <i>r</i> is biased towards the smallest crystal dimension. The results presented here suggest that the IUCr publication requirement that a numerical absorption correction must be applied for strongly absorbing crystals is in need of revision.
Ähnliche Arbeiten
Handbook of X-Ray Photoelectron Spectroscopy
1995 · 23.683 Zit.
<i>ATHENA</i>,<i>ARTEMIS</i>,<i>HEPHAESTUS</i>: data analysis for X-ray absorption spectroscopy using<i>IFEFFIT</i>
2005 · 16.045 Zit.
Resolving surface chemical states in XPS analysis of first row transition metals, oxides and hydroxides: Cr, Mn, Fe, Co and Ni
2010 · 10.328 Zit.
X-ray line broadening from filed aluminium and wolfram
1953 · 10.012 Zit.
High-Resolution X-Ray Photoemission Spectrum of the Valence Bands of Gold
1972 · 6.110 Zit.