OpenAlex · Aktualisierung stündlich · Letzte Aktualisierung: 30.03.2026, 13:58

Dies ist eine Übersichtsseite mit Metadaten zu dieser wissenschaftlichen Arbeit. Der vollständige Artikel ist beim Verlag verfügbar.

Methodological assessment of the reduction of dissemination risk and quantification of debris dispersion during dissection with a surgical aspirator

2022·2 Zitationen·BMC Research NotesOpen Access
Volltext beim Verlag öffnen

2

Zitationen

12

Autoren

2022

Jahr

Abstract

Methylene blue-stained brain phantom gelatin was immersed in a chamber with distilled water solution and dissected with an ADPJ. The dispersed gelatin solution was stirred and warmed to dissolve. The absorbance of the solution was measured spectrophotometrically. First, a reference standard curve was constructed to confirm the relationship between the absorbance and the amount of the dispersed gelatin. A clear proportional correlation was observed, which indicated that absorbance measurements can help evaluate the amount of dispersion. Using this method, we revealed that a high dissection force, insufficient suction, and inappropriate long distance between the nozzle tip and the target were associated with increased dispersion. This method might constitute a versatile and reliable approach to evaluate dispersion and aid in the development of surgical devices.